Choose the right accessories for your application

Different Andrew accessories are available to better integrate into your laboratory workflow

CaliBro

CaliBro

Andrew CaliBro is a revolutionary suite designed around Andrew, which allows any laboratory to check the calibration of a set of up to 5 pipettes with a simple click of a button.

With its totally unattended and effective operation, Andrew CaliBro validates the performance of each selected pipette and its compliance to the specifications, while in the end of the validation provides an individual validation certificate for each pipette. It is an effective Operational Qualification (OQ) of pipettes and Andrew itself.

Andrew CaliBro can:

TECHNICAL SPECIFICATIONS

PLATE TYPES
96 well microplates with flat, round or UV bottom

WAVELENGTH RANGE
340 – 900 nm

MEASUREMENT RANGE
0.000 – 4.000 OD (Abs)

OPTICAL SYSTEM
Maintenance free 8 channel optical system, self-calibrating, no reference channel required

LIGHT SOURCE
LED, wavelength specific (life time = 1000 x Halogen Lamp)

DETECTORS
Photodiodes

USED WAVELENGTH(S)
1 wavelength 520nm

PC-INTERFACE
USB 2.0

PC SOFTWARE
AndrewOS and CaliBro software (included)

POWER REQUIREMENTS
100 – 240V, 50 or 60Hz (autosensing)

POWER CONSUMPTION
Operational: max 20w
Standby: max 2W

MEASUREMENT SYSTEM
– Maintenance Free due to capsulated optical cylinders
– Light sources: LED (life time = 1000 x Halogen Lamp)
– 8-channel fiber optics
– Detectors:Photodiodes
– Wavelength range: 340 – 900 nm

MEASURES
Instrument only (net):
– Dimensions: 23 x 11,5 x 36 cm (w x h x d)
– weight: 6.4 kg (14.1 lbs)
Instrument packed incl. accessories (gross):
– Dimensions: 31 x 36 x 48 cm (w x h x d)
– weight: 8,3 kg (18,3 lbs)

ENVIRONMENTAL
Suggested operational environment:
23±3 °C 50 – 90% rel. Hum. (non condensing)
Storage :
20°C – 35°C
5 – 95% rel. Hum. (non condensing)

OPTIONS
other light cylinders on request

OVERVOLTAGE CAT.:
II

POLLUTION DEGREE:
2

METHOD OF DISPOSAL:
Contaminated Waste

USAGE:
Commercial

Description